TWO-DEVICE ENERGY WORKSHEET · OCTOBER 2026

Plan two CPAP machines as two separate medical-device loads.

Two people can use different machines, pressure settings, humidification, tubing, masks, and power paths. Measure each complete overnight configuration separately, then combine the energy only after preserving each device's instructions and continuity plan.

This is not medical guidance or a compatibility approvalDo not rely on this worksheet for life-sustaining equipment. Follow each device manufacturer's instructions, contact the equipment supplier, and discuss outage continuity with the appropriate healthcare professionals. If loss of therapy creates immediate danger, use the emergency plan and seek emergency services.

Affiliate disclosure: Power Prep Labs may eventually earn a commission from clearly labeled shopping links, at no extra cost to you. No paid links are currently active. This worksheet does not approve any station for medical use.

Measure first; do not double an internet average

Record cumulative watt-hours for one representative night on each exact machine using the intended humidifier, heated tube, mask, pressure settings, and manufacturer-permitted power path. A power-supply label or maximum rating is not the same as overnight energy. Repeat if seasonal humidity settings or room conditions materially change.

Combine two measured nights

The prefilled values—300 Wh and 450 Wh per night—are arithmetic examples only, not CPAP consumption claims. Replace them with separate measurements for the two actual configurations.

Combined energy/night750 Wh
Raw outage energy1,500 Wh
Rated capacity needed*1,961 Wh
Planning classAt least 2 kWh

* Energy calculation only. It does not establish manufacturer permission, connector compatibility, transfer continuity, power quality, output capacity, alarm behavior, recharge time, or clinical adequacy. Round up to an exact product capacity only after every separate requirement is verified.

Worked example

combined nightly energy = device A Wh + device B Whraw outage energy = combined nightly Wh × nights + other Whrated capacity needed = raw Wh ÷ efficiency ÷ (1 − reserve)

For the illustrative values, (300 + 450) × 2 = 1,500 Wh. At 85% conversion efficiency with 10% held in reserve, the rated-capacity result is 1,500 ÷ 0.85 ÷ 0.90 = 1,961 Wh. A rounded “2 kWh” label is not enough evidence: compare the exact capacity and every device-specific requirement.

Why the two measurements may differ

ResMed's AirSense 11 clinical guide, for example, lists typical and peak power consumption for the system under its stated conditions, while its user guide identifies a 65 W adapter and compatible accessories. Philips' DreamStation 2 instructions separately note that some battery operation does not support humidification and that an 80 W supply is required to support humidification. These are exact-family instructions—not interchangeable inputs for another machine.

One station creates a common failure point

If two users share one battery, inverter, cable path, or recharge plan, one fault can interrupt both devices. Consider whether the continuity plan needs separate approved batteries, a second station, spare adapters, an alternate location with power, or another clinically appropriate fallback. The correct redundancy depends on the consequences of interruption and must be decided with the equipment suppliers and care professionals.

Transfer time is not assumed safe

A portable station marketed with UPS or EPS behavior may still interrupt power for a stated number of milliseconds, impose a lower bypass limit, restart differently after depletion, or be prohibited for the device. Do not infer zero-interruption medical suitability from a marketing label. Verify each device's instructions and test only within manufacturer and clinical guidance.

Recharging is part of the capacity decision

Two nights of stored energy do not create a third night. Document how the battery will recharge during the same weather that caused the outage, how long the exact recharge takes, whether that source remains accessible, and what happens if charging fails. Keep all equipment dry and inspect cords and batteries for damage.

Build the owner-specific continuity sheet

Record separately for each personWhy it matters
Exact device, serial/model family, supplier, and clinical contactInstructions, approved accessories, and urgency are user-specific
Measured Wh for the exact overnight configurationPrevents one user's energy estimate from being copied to the other
Approved AC/DC adapter and spareConnector, voltage, polarity, and permitted power path can differ
Interruption tolerance and emergency actionA calculator cannot decide the consequence of lost therapy
Backup source, recharge source, and responsible personTurns stored energy into an actionable continuity plan
Last successful test and battery-maintenance dateConfirms the plan is maintained rather than merely purchased

Direct answers

What size battery for two CPAP machines?

Add separate measured watt-hours per night, multiply by the planned nights, include other loads, and divide by conversion efficiency and usable battery share. The result estimates energy only; use exact manufacturer instructions to decide whether the power path is permitted.

Can both machines share one station?

Only after both manufacturers' requirements and the exact station's energy, output, ports, waveform, transfer, and failure behavior are verified. Shared hardware also concentrates risk.

Should humidification be disabled to extend runtime?

It can change energy materially, but do not alter a therapy configuration based on a generic runtime calculation. Follow the device instructions and healthcare guidance.

Primary sources and next steps